، جلد ۸، شماره ۲، صفحات ۸۲-۸۵

عنوان فارسی
چکیده فارسی مقاله
کلیدواژه‌های فارسی مقاله

عنوان انگلیسی Missing Part in Adjustable Pressure Limiting Valve in Bain’s Circuit: A Near Miss
چکیده انگلیسی مقاله Breathing circuits are the delivery systems which conduct anaesthetic gases from an anaesthetic machine to the patient. They are designed for either spontaneous respiration or intermittent positive pressure ventilation (IPPV). The Bain co-axial circuit, a modification of Mapleson D system is the most commonly used circuit in outside operating room procedures. It has several advantages over the circle system in that it is light-weight, convenient with adjustable length. Although most problems with the circuit involve disconnections, obstruction problems related to kinks in the gas delivery tube  or foreign body in the elbow have been reported. We would like to report an anaesthetic incident involving desaturation of patient posted for electro convulsive therapy (ECT) due to missing part of the Adjustable Pressure Limiting (APL) valve.
کلیدواژه‌های انگلیسی مقاله

نویسندگان مقاله | Divya Devanathan
Department of Anesthesiology, Mahatma Gandhi Medical College, Sri Balaji Vidyapeeth University, Pondicherry.



نشانی اینترنتی https://crcp.tums.ac.ir/index.php/crcp/article/view/777
فایل مقاله فایلی برای مقاله ذخیره نشده است
کد مقاله (doi)
زبان مقاله منتشر شده
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده
برگشت به: صفحه اول پایگاه   |   نسخه مرتبط   |   نشریه مرتبط   |   فهرست نشریات